IF5304 (Discontinued) 1700C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace KSL-1700X-G216

Catalog No. IF5304
Item Number KSL-1700X-G216
Working Temperature 1600℃ continuously
Max Working Temp. (℃) 1700℃, ≤1hour
Chamber Size 600*600*600mm (216L)

Discontinued

1700C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace KSL-1700X-G216 is a bottom loading larger box furnace (216 Liter capacity) with inert gas atmosphere control up to 1700℃. Stanford Advanced Materials (SAM) boasts extensive experience in providing Muffle Furnaces with both high purity and competitive pricing.

Related products: 1100C Compact Hybrid Furnace KSL-1100X-S-H, 1800C Bench-Top Muffle Furnace KSL-1800X-A1-W

INQUIRY
Add to Inquiry List
24-ksl-1700x-g216
24-ksl-1700x-g216
24-ksl-1700x-g216
24-ksl-1700x-g216
Description
Specification

1700C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace KSL-1700X-G216 Description

1700C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace KSL-1700X-G216 is a larger bottom-loading box furnace with a capacity of 216 liters, offering inert gas atmosphere control up to 1700℃. Featuring high-quality alumina fiber bricks and Molybdenum alloy heating elements, the furnace includes a 24"x24"x24" heating chamber. Its vacuum-sealed steel chamber is equipped with a water-cooling jacket, and the vacuum and gas flow control system facilitates material processing under an inert gas atmosphere.

1700C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace KSL-1700X-G216 Specifications

Furnace Structure

  • Double-layer stainless steel structure for maximum safety
  • The bottom loading mechanism adopts electric hydraulic lifting which can carry up to 200 Kg of sample
  • Water cooling front door and sealing jacket for longer service life.
  • Touch-screen digital control panel for all parameters.
  • Precision temperature control with +/- 3℃ uniformity inside the chamber and +/-1℃ accuracy.

Inside Chamber Size

  • Heated chamber: 600 x 600 x 600 mm (24" x 24" x 24". 216 Liter capacity)
  • Energy-saving 1800-grade fibrous alumina insulation.
  • High-purity Al2O3 coating.

Power 

  • AC 460V, three-phase; 50/60 Hz
  • 50 KW
  • 63Amp (100Aair-breaker)

Heating Element

12 Pcs Mo heating element

Continuous Working Temperature

1600℃

Maximum Working Temperature

1700℃ (< 1 hr.)

Heating Rate

5℃ / min

Temperature Accuracy

+/- 0.5℃

Temperature Control

+/ - 1℃ during the dwelling.

Thermocouple

two B-type thermocouples (control thermocouple and secondary alarm thermocouple)

Temperature Control

  • Eurotherm controller with PID automatic control via SCR (Silicon Controlled Rectifier)
  • 30 programmable segments for precise control of heating/cooling rate and dwell time.
  • Over-temperature protection and thermocouple failure alarm functions allow operation without attendant(s).
  • +/- 0.5℃ temperature accuracy.
  • RS485 Communications Port

Gas Flow Control and Vacuum Valve

  • Two flow meters (#1: 0~8L/min; #2: 0~4L/min) are installed on the front panel of the furnace.
  • One heavy-duty vacuum pump (CE certified) is built on the bottom of the furnace.

Water Cooling

  • A 16L/min recirculating water chiller is installed in the bottom of the furnace for cooling the furnace sealing flange (CE certified only)
  • MET Certification (UL 1450) for water chiller is available upon request at extra cost, please contact our sales representatives for a quote.

Net Weight

3000 Kg

Warranty

One-year limited warranty for electrical and mechanical parts 

Compliance

  • CE Certified for the furnace, vacuum pump, and water chiller.
  • NRTL (UL61010) or CSA certification is available at extra cost.

Application Notes

  • For longer furnace life furnaces, we recommend the heating rate not to exceed 5 ℃ /min.
  • The cooling rate shall also not exceed 5 ℃ /min.
  • Fine cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with alumina coating.
  • The refractory block must be inserted before closing the door.
  • Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation.

Warning

  • ATTENTION: Any damages caused by the use of corrosive and acidic gasses are not under the coverage of SAM One Year Limited Warranty.
  • To prevent oxidation/corrosion damages to the Mo heating elements, it is necessary to keep a constant inert gas (Ar preferred) flow throughout the entire process. This is the most effective way to carry out any by-products.

1700C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace KSL-1700X-G216 Applications

1700C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace KSL-1700X-G216 is suitable for industrial or laboratory sintering of larger materials.

1700C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace KSL-1700X-G216 Packaging

Our 1700C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace KSL-1700X-G216 is carefully handled during storage and transportation to preserve the quality of our product in its original condition.

GET A QUOTE

Send us an Inquiry now to find out more Information and the latest prices,thanks!

* Your Name
* Your Email
* Product name
* Your Phone
* Country

United States

    Comments
    * Check Code
    Leave A Message
    Leave A Message
    * Your Name:
    * Your Email:
    * Product name:
    * Your Phone:
    * Comments: